RIGOL Chapter 2 Command System
2-40 DL3000 Programming Guide
[:SOURce]:BATTary:TENabstop
Syntax [:SOURce]:BATTary:TENabstop {0|1|ON|OFF}
[:SOURce]:BATTary:TENabstop?
Description Sets the discharge cut-off time switch of the electronic load.
Queries the on/off status of the discharge cut-off time of the electronic load.
Return
Format
0 (OFF) or 1 (ON)
Example
:SOUR:BATT:TEN 1 /*Enables the discharge cut-off time switch of the
electronic load.*/
:SOUR:BATT:TEN? /*Queries the on/off status of the discharge cut-off time of
the electronic load.*/
[:SOURce]:OCP:RANGe
Syntax [:SOURce]:OCP:RANGe {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OCP:RANGe? [MINimum|MAXimum|DEFault]
Description Sets the current range in OCP mode to be a high range or a low one.
Queries the current range set in OCP mode.
Remarks The default range is a high range.
In the setting command, the parameter "MINimum" indicates setting a low range;
the parameter "MAXimum" indicates setting a high range; and the parameter
"DEFault" indicates setting the default range.
In the query command, the parameter "MINimum" indicates querying a low range;
the parameter "MAXimum" indicates querying a high range; and the parameter
"DEFault" indicates querying the default range.
Return
Format
The query returns a real number.
Example
:SOUR:OCP:RANG 60 /*Sets the current range in OCP mode to be 60 A (a high
range).*/
:SOUR:OCP:RANG? /*Queries the current range set in OCP mode.*/
[:SOURce]:OCP:VON
Syntax [:SOURce]:OCP:VON {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OCP:VON? [MINimum|MAXimum|DEFault]
Description Sets the starting voltage in OCP mode.
Queries the starting voltage in OCP mode.
Remarks The default unit for the starting voltage is V, and it ranges from 0 V to 150 V.
In the setting command, the parameter "MINimum" indicates setting the minimum
starting voltage; the parameter "MAXimum" indicates setting the maximum starting
voltage; and the parameter "DEFault" indicates setting the default starting voltage.
In the query command, the parameter "MINimum" indicates querying the minimum
starting voltage; the parameter "MAXimum" indicates querying the maximum
starting voltage; and the parameter "DEFault" indicates querying the default starting
voltage.

Table of Contents

Other manuals for Rigol DL3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DL3000 and is the answer not in the manual?

Rigol DL3000 Specifications

General IconGeneral
BrandRigol
ModelDL3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals