Chapter 2 Command System RIGOL
DL3000 Programming Guide 2-39
[:SOURce]:BATTary:VON
Syntax [:SOURce]: BATTary:VON {<value>|MINimum|MAXimum|DEFault}
[:SOURce]: BATTary:VON? [MINimum|MAXimum|DEFault]
Description Sets the starting voltage in Battery mode.
Queries the starting voltage in Battery mode.
Remarks The default unit for the starting voltage is V, and it ranges from 0 V to 150 V.
In the setting command, the parameter "MINimum" indicates setting the minimum
starting voltage; the parameter "MAXimum" indicates setting the maximum starting
voltage; and the parameter "DEFault" indicates setting the default starting voltage.
In the query command, the parameter "MINimum" indicates querying the minimum
starting voltage; the parameter "MAXimum" indicates querying the maximum
starting voltage; and the parameter "DEFault" indicates querying the default starting
Return
The query returns a real number.
Example
:SOUR:BATT:VON 1 /*Sets the starting voltage in Battery mode to be 1 V.*/
:SOUR:BATT:VON? /*Queries the starting voltage in Battery mode.*/
[:SOURce]:BATTary:VENabstop
Syntax [:SOURce]:BATTary:VENabstop {0|1|ON|OFF}
[:SOURce]:BATTary:VENabstop?
Description Sets the cut-off voltage switch of the electronic load.
Queries the on/off status of the cut-off voltage of the electronic load.
Return
0 (OFF) or 1 (ON)
Example
:SOUR:BATT:VEN 1 /*Enables the cut-off voltage of the electronic load.*/
:SOUR:BATT:VEN? /*Queries the on/off status of the cut-off voltage of the
[:SOURce]:BATTary:CENabstop
Syntax [:SOURce]:BATTary:CENabstop {0|1|ON|OFF}
[:SOURce]:BATTary:CENabstop?
Description Sets the battery capacity switch of the discharge cutoff of the electronic load.
Queries the battery capacity on/off status of the discharge cutoff of the electronic load.
Return
0 (OFF) or 1 (ON)
Example
:SOUR:BATT:CEN 1 /*Enables the battery capacity of the discharge cutoff of
the electronic load.*/
:SOUR:BATT:CEN? /*Queries the battery capacity on/off status of the
discharge cutoff of the electronic load.*/