Chapter 2 Command System RIGOL
DL3000 Programming Guide 2-43
OCP mode. The default unit for the value is A.
In the setting command, the parameter "MINimum" indicates setting the minimum
value of the maximum current; the parameter "MAXimum" indicates setting the
maximum value of the maximum current; and the parameter "DEFault" indicates
setting the default value of the maximum current.
In the query command, the parameter "MINimum" indicates querying the minimum
value of the maximum current; the parameter "MAXimum" indicates querying the
maximum value of the maximum current; and the parameter "DEFault" indicates
querying the default value of the maximum current.
Return
Format
The query returns a real number.
Example
:SOUR:OCP:IMAX 10 /*Sets the maximum current in OCP mode to be 10 A.*/
:SOUR:OCP:IMAX? /*Queries the maximum current in OCP mode.*/
[:SOURce]:OCP:IMIN
Syntax [:SOURce]:OCP:IMIN {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OCP:IMIN? [MINimum|MAXimum|DEFault]
Description Sets the minimum current in OCP mode.
Queries the minimum current in OCP mode.
Remarks The minimum current is called the minimum value of the protection current in the
OCP mode. The default unit for the value is A.
In the setting command, the parameter "MINimum" indicates setting the minimum
value of the minimum current; the parameter "MAXimum" indicates setting the
maximum value of the minimum current; and the parameter "DEFault" indicates
setting the default value of the minimum current.
In the query command, the parameter "MINimum" indicates querying the minimum
value of the minimum current; the parameter "MAXimum" indicates querying the
maximum value of the minimum current; and the parameter "DEFault" indicates
querying the default value of the minimum current.
Return
Format
The query returns a real number.
Example
:SOUR:OCP:IMIN 1 /*Sets the minimum current in OCP mode to be 1 A.*/
:SOUR:OCP:IMIN? /*Queries the minimum current in OCP mode.*/
[:SOURce]:OCP:VOCP
Syntax [:SOURce]:OCP:VOCP {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OCP:VOCP? [MINimum|MAXimum|DEFault]
Description Sets the protection voltage in OCP mode.
Queries the protection voltage in OCP mode.
Remarks The default unit for the protection voltage is V.
In the setting command, the parameter "MINimum" indicates setting the minimum
protection voltage; the parameter "MAXimum" indicates setting the maximum
protection voltage; and the parameter "DEFault" indicates setting the default
protection voltage.

Table of Contents

Other manuals for Rigol DL3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DL3000 and is the answer not in the manual?

Rigol DL3000 Specifications

General IconGeneral
BrandRigol
ModelDL3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals