Chapter 2 Command System      RIGOL 
DL3000 Programming Guide  2-27 
Return 
Format 
The query returns a real number. 
Example 
:SOUR:CURR:TRAN:AWID 1   /*Sets the width of Level A in continuous or pulsed 
transient operation to be 1 ms.*/ 
:SOUR:CURR:TRAN:AWID?   /*Queries the width of Level A in continuous or pulsed 
Related 
Command 
 
 
[:SOURce]:CURRent:TRANsient:BWIDth 
 
 
[:SOURce]:CURRent:TRANsient:BWIDth 
Syntax  [:SOURce]:CURRent:TRANsient:BWIDth {<value>|MINimum|MAXimum|DEFault} 
[:SOURce]:CURRent:TRANsient:BWIDth? [MINimum|MAXimum|DEFault] 
Description  Sets the width of Level B in continuous and pulsed transient operation. 
Sets the width of Level B in continuous and pulsed transient operation. 
Remarks    The width of Level B refers to the time during which the sink current stays at Level B 
when it switches to Level B in continuous transient operation mode. 
  In the setting command, the parameter "MINimum" indicates setting the minimum 
width of Level B; the parameter "MAXimum" indicates setting the maximum width of 
Level B; and the parameter "DEFault" indicates setting the default width of Level B. 
  In the query command, the parameter "MINimum" indicates querying the minimum 
width of Level B; the parameter "MAXimum" indicates querying the maximum width 
of Level B; and the parameter "DEFault" indicates querying the default width of Level 
Return 
The query returns a real number. 
Example 
:SOUR:CURR:TRAN:BWID 1   /*Sets the width of Level B in continuous or pulsed 
transient operation to be 1 ms.*/ 
:SOUR:CURR:TRAN:BWID?    /*Queries the width of Level B in continuous or pulsed 
Related 
[:SOURce]:CURRent:TRANsient:AWIDth 
 
 
[:SOURce]:CURRent:TRANsient:FREQuency 
Syntax  [:SOURce]:CURRent:TRANsient:FREQuency {<value>|MINimum|MAXimum|DEFault} 
[:SOURce]:CURRent:TRANsient:FREQuency? [MINimum|MAXimum|DEFault] 
Description  Sets the frequency in continuous mode. 
Queries the frequency set in continuous mode. 
Remarks    Frequency is the reciprocal of period. The unit of frequency is kHz. 
  In the setting command, the parameter "MINimum" indicates setting minimum 
frequency; the parameter "MAXimum" indicates setting maximum frequency; and the 
parameter "DEFault" indicates setting the default frequency. 
  In the query command, the parameter "MINimum" indicates querying the minimum