Chapter 2 Command System RIGOL
DL3000 Programming Guide 2-27
Return
Format
The query returns a real number.
Example
:SOUR:CURR:TRAN:AWID 1 /*Sets the width of Level A in continuous or pulsed
transient operation to be 1 ms.*/
:SOUR:CURR:TRAN:AWID? /*Queries the width of Level A in continuous or pulsed
transient operation.*/
Related
Command
[:SOURce]:CURRent:TRANsient:BWIDth
[:SOURce]:CURRent:TRANsient:BWIDth
Syntax [:SOURce]:CURRent:TRANsient:BWIDth {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:CURRent:TRANsient:BWIDth? [MINimum|MAXimum|DEFault]
Description Sets the width of Level B in continuous and pulsed transient operation.
Sets the width of Level B in continuous and pulsed transient operation.
Remarks The width of Level B refers to the time during which the sink current stays at Level B
when it switches to Level B in continuous transient operation mode.
In the setting command, the parameter "MINimum" indicates setting the minimum
width of Level B; the parameter "MAXimum" indicates setting the maximum width of
Level B; and the parameter "DEFault" indicates setting the default width of Level B.
In the query command, the parameter "MINimum" indicates querying the minimum
width of Level B; the parameter "MAXimum" indicates querying the maximum width
of Level B; and the parameter "DEFault" indicates querying the default width of Level
B.
Return
Format
The query returns a real number.
Example
:SOUR:CURR:TRAN:BWID 1 /*Sets the width of Level B in continuous or pulsed
transient operation to be 1 ms.*/
:SOUR:CURR:TRAN:BWID? /*Queries the width of Level B in continuous or pulsed
transient operation.*/
Related
Command
[:SOURce]:CURRent:TRANsient:AWIDth
[:SOURce]:CURRent:TRANsient:FREQuency
Syntax [:SOURce]:CURRent:TRANsient:FREQuency {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:CURRent:TRANsient:FREQuency? [MINimum|MAXimum|DEFault]
Description Sets the frequency in continuous mode.
Queries the frequency set in continuous mode.
Remarks Frequency is the reciprocal of period. The unit of frequency is kHz.
In the setting command, the parameter "MINimum" indicates setting minimum
frequency; the parameter "MAXimum" indicates setting maximum frequency; and the
parameter "DEFault" indicates setting the default frequency.
In the query command, the parameter "MINimum" indicates querying the minimum

Table of Contents

Other manuals for Rigol DL3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DL3000 and is the answer not in the manual?

Rigol DL3000 Specifications

General IconGeneral
BrandRigol
ModelDL3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals