RIGOL Chapter 2 Command System
2-46 DL3000 Programming Guide
[:SOURce]:OPP:PSTep
Syntax [:SOURce]:OPP:PSTep {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OPP:PSTep? [MINimum|MAXimum|DEFault]
Description Sets the step power in OPP mode.
Queries the step power in OPP mode.
Remarks The step power is called the step value of the power in OP
P mode. The default unit
is W.
In the setting command, the parameter "MINimum" indicates setting the minimum
step power; the parameter "MAXimum" indicates setting the maximum step power;
and the parameter "DEFault" indicates setting the default step power.
In the query command, the parameter "MINimum" indicates querying the minimum
step power; the parameter "MAXimum" indicates querying the maximum step
power; and the parameter "DEFault" indicates querying the default step power.
Return
Format
The query returns a real number.
Example
:SOUR:OPP:PST 1 /*Sets the step power in OPP mode to be 1 W.*/
:SOUR:OPP:PST? /*Queries the step power in OPP mode.*/
[:SOURce]:OPP:PDELaystep
Syntax [:SOURce]:OPP:PDELaystep {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OPP:PDELaystep? [MINimum|MAXimum|DEFault]
Description Sets the delay time of the step power in OPP mode.
Queries the delay time of the step power in OPP mode.
Remarks The default unit for the delay time is ms.
In the setting command, the parameter "MINimum" indicates setting the minimum
delay time; the parameter "MAXimum" indicates setting the maximum delay time;
and the parameter "DEFault" indicates setting the default delay time.
In the query command, the parameter "MINimum" indicates querying the minimum
delay time; the parameter "MAXimum" indicates querying the maximum delay time;
and the parameter "DEFault" indicates querying the default delay time.
Return
Format
The query returns a real number.
Example
:SOUR:OPP:PDEL 500 /*Sets the delay time of the step power in OPP mode to be
500 ms.*/
:SOUR:OPP:PDEL? /*Queries the delay time of the step power in OPP mode.*/
[:SOURce]:OPP:PMAX
Syntax [:SOURce]:OPP:PMAX {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OPP:PMAX? [MINimum|MAXimum|DEFault]
Description Sets the maximum power in OPP mode.
Queries the maximum power in OPP mode.
Remarks
The maximum power is called the maximum value of the protection power in the OPP

Table of Contents

Other manuals for Rigol DL3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DL3000 and is the answer not in the manual?

Rigol DL3000 Specifications

General IconGeneral
BrandRigol
ModelDL3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals