Chapter 2 Command System RIGOL
DL3000 Programming Guide 2-45
Return
Format
The query returns a real number.
Example
:SOUR:OPP:VON 5 /*Sets the starting voltage in OPP mode to be 5 V.*/
:SOUR:OPP:VON? /*Queries the starting voltage in OPP mode.*/
[:SOURce]:OPP:VONDelay
Syntax [:SOURce]:OPP:VONDelay {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OPP:VONDelay? [MINimum|MAXimum|DEFault]
Description Sets the delay time of the starting voltage in OPP mode.
Queries the delay time of the starting voltage in OPP mode.
Remarks When the input voltage reaches the starting voltage and delays for a period of time,
the load starts to sink the current.
The default unit for the delay time is ms.
In the setting command, the parameter "MINimum" indicates setting the minimum
delay time; the parameter "MAXimum" indicates setting the maximum delay time;
and the parameter "DEFault" indicates setting the default delay time.
In the query command, the parameter "MINimum" indicates querying the minimum
delay time; the parameter "MAXimum" indicates querying the maximum delay time;
and the parameter "DEFault" indicates querying the default delay time.
Return
Format
The query returns a real number.
Example
:SOUR:OPP:VOND 500 /*Sets the delay time of the starting voltage in OPP mode
to be 500 ms.*/
:SOUR:OPP:VOND? /*Queries the delay time of the starting voltage in OPP
mode.*/
[:SOURce]:OPP:PSET
Syntax [:SOURce]:OPP:PSET {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:OPP:PSET? [MINimum|MAXimum|DEFault]
Description Sets the starting power in OPP mode.
Queries the starting power in OPP mode.
Remarks The starting power is called the power value when the load starts in OPP mode. The
default unit for the value is W.
In the setting command, the parameter "MINimum" indicates setting the minimum
starting power; the parameter "MAXimum" indicates setting the maximum starting
power; and the parameter "DEFault" indicates setting the default starting power.
In the query command, the parameter "MINimum" indicates querying the minimum
starting power; the parameter "MAXimum" indicates querying the maximum starting
power; and the parameter "DEFault" indicates querying the default starting power.
Return
Format
The query returns a real number.
Example
:SOUR:OPP:PSET 5 /*Sets the starting power in OPP mode to be 5 W.*/
:SOUR:OPP:PSET? /*Queries the starting power in OPP mode.*/

Table of Contents

Other manuals for Rigol DL3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DL3000 and is the answer not in the manual?

Rigol DL3000 Specifications

General IconGeneral
BrandRigol
ModelDL3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals