RIGOL  Chapter 2 Command System 
2-26  DL3000 Programming Guide 
parameter "DEFault" indicates querying the default value of Level A. 
Return 
The query returns a real number. 
Example 
:SOUR:CURR:TRAN:ALEV 5      /*Sets Level A in transient operation mode to be 5 A.*/ 
:SOUR:CURR:TRAN:ALEV?      /*Queries Level A set in transient operation mode.*/ 
Related 
[:SOURce]:CURRent:TRANsient:BLEVel 
 
 
[:SOURce]:CURRent:TRANsient:BLEVel 
Syntax  [:SOURce]:CURRent:TRANsient:BLEVel {<value>|MINimum|MAXimum|DEFault} 
[:SOURce]:CURRent:TRANsient:BLEVel? [MINimum|MAXimum|DEFault] 
Description  Sets Level B in transient operation mode. 
Queries Level B set in transient operation mode. 
Remarks    In transient operation, the sink current toggles between a high value and a low value. 
Level B indicates a low value. The unit for Level B is Amphere (A). 
  The input value of Level B should be within the set range. 
  In the setting command, the parameter "MINimum" indicates setting the minimum 
Level B; the parameter "MAXimum" indicates setting the maximum Level B; and the 
parameter "DEFault" indicates setting the default value of Level B. 
  In the query command, the parameter "MINimum" indicates querying the minimum 
Level B; the parameter "MAXimum" indicates querying the maximum Level B; and the 
parameter "DEFault" indicates querying the default value of Level B.   
Return 
The query returns a real number. 
Example 
:SOUR:CURR:TRAN:BLEV 5       /*Sets Level B in transient operation mode to be 5 A.*/ 
:SOUR:CURR:TRAN:BLEV?        /*Queries Level B set in transient operation mode.*/ 
Related 
[:SOURce]:CURRent:TRANsient:ALEVel 
 
 
[:SOURce]:CURRent:TRANsient:AWIDth 
Syntax  [:SOURce]:CURRent:TRANsient:AWIDth {<value>|MINimum|MAXimum|DEFault} 
[:SOURce]:CURRent:TRANsient:AWIDth? [MINimum|MAXimum|DEFault] 
Description  Sets the width of Level A in continuous and pulsed transient operation. 
Sets the width of Level A in continuous and pulsed transient operation. 
Remarks    The width of Level A refers to the time during which the sink current stays at Level A 
when it switches to Level A in continuous and pulsed transient operation mode.   
  In the setting command, the parameter "MINimum" indicates setting the minimum 
width of Level A; the parameter "MAXimum" indicates setting the maximum width of 
Level A; and the parameter "DEFault" indicates setting the default width of Level A. 
  In the query command, the parameter "MINimum" indicates querying the minimum 
width of Level A; the parameter "MAXimum" indicates querying the maximum width 
of Level A; and the parameter "DEFault" indicates querying the default width of Level