RIGOL Chapter 2 Command System
2-26 DL3000 Programming Guide
parameter "DEFault" indicates querying the default value of Level A.
Return
The query returns a real number.
Example
:SOUR:CURR:TRAN:ALEV 5 /*Sets Level A in transient operation mode to be 5 A.*/
:SOUR:CURR:TRAN:ALEV? /*Queries Level A set in transient operation mode.*/
Related
[:SOURce]:CURRent:TRANsient:BLEVel
[:SOURce]:CURRent:TRANsient:BLEVel
Syntax [:SOURce]:CURRent:TRANsient:BLEVel {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:CURRent:TRANsient:BLEVel? [MINimum|MAXimum|DEFault]
Description Sets Level B in transient operation mode.
Queries Level B set in transient operation mode.
Remarks In transient operation, the sink current toggles between a high value and a low value.
Level B indicates a low value. The unit for Level B is Amphere (A).
The input value of Level B should be within the set range.
In the setting command, the parameter "MINimum" indicates setting the minimum
Level B; the parameter "MAXimum" indicates setting the maximum Level B; and the
parameter "DEFault" indicates setting the default value of Level B.
In the query command, the parameter "MINimum" indicates querying the minimum
Level B; the parameter "MAXimum" indicates querying the maximum Level B; and the
parameter "DEFault" indicates querying the default value of Level B.
Return
The query returns a real number.
Example
:SOUR:CURR:TRAN:BLEV 5 /*Sets Level B in transient operation mode to be 5 A.*/
:SOUR:CURR:TRAN:BLEV? /*Queries Level B set in transient operation mode.*/
Related
[:SOURce]:CURRent:TRANsient:ALEVel
[:SOURce]:CURRent:TRANsient:AWIDth
Syntax [:SOURce]:CURRent:TRANsient:AWIDth {<value>|MINimum|MAXimum|DEFault}
[:SOURce]:CURRent:TRANsient:AWIDth? [MINimum|MAXimum|DEFault]
Description Sets the width of Level A in continuous and pulsed transient operation.
Sets the width of Level A in continuous and pulsed transient operation.
Remarks The width of Level A refers to the time during which the sink current stays at Level A
when it switches to Level A in continuous and pulsed transient operation mode.
In the setting command, the parameter "MINimum" indicates setting the minimum
width of Level A; the parameter "MAXimum" indicates setting the maximum width of
Level A; and the parameter "DEFault" indicates setting the default width of Level A.
In the query command, the parameter "MINimum" indicates querying the minimum
width of Level A; the parameter "MAXimum" indicates querying the maximum width
of Level A; and the parameter "DEFault" indicates querying the default width of Level