EasyManuals Logo

Rigol DS1054Z Programming Guide

Rigol DS1054Z
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-23
:CURSor:MANual:IXDELta?
Syntax
:CURSor:MANual:IXDELta?
Description
Query the reciprocal of the absolute value of the difference between the X values of cursor
A and cursor B (1/|dX|) in the manual cursor measurement mode. The unit depends on
the horizontal unit currently selected.
Return
Format
The query returns 1/|dX| in scientific notation.
Example
:CURSor:MANual:IXDELta? /*The query returns 1.250000e+05*/
Related
Commands
:CURSor:MANual:AX
:CURSor:MANual:BX
:CURSor:MANual:TUNit
:CURSor:MANual:YDELta?
Syntax
:CURSor:MANual:YDELta?
Description
Query the difference between the Y values of cursor A and cursor B (BY-AY) in the manual
cursor measurement mode. The unit depends on the vertical unit currently selected.
Return
Format
When the signal source is CHANnel1|CHANnel2|CHANnel3|CHANnel4|MATH, the query
returns the difference in scientific notation; when the signal source is LA, the query
returns 4294967295.
Example
:CURSor:MANual:YDELta? /*The query returns -4.000000e+00*/
Related
Commands
:CURSor:MANual:AY
:CURSor:MANual:BY
:CURSor:MANual:VUNit

Table of Contents

Other manuals for Rigol DS1054Z

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1054Z and is the answer not in the manual?

Rigol DS1054Z Specifications

General IconGeneral
Channels4
Bandwidth50 MHz
Sample Rate1 GSa/s
Memory Depth24 Mpts
Display7 inch TFT LCD (800x480)
Vertical Sensitivity1 mV/div to 10 V/div
Time Base Range5 ns/div to 50 s/div
Vertical Resolution8 bits
Trigger TypesEdge, Pulse, Video, Slope
ConnectivityUSB Host, USB Device, LAN
Input Impedance1 MΩ
Power Supply100-240 VAC, 45-440 Hz