RIGOL Chapter 2 Command System
2-64 MSO1000Z/DS1000Z Programming Guide
:DISPlay:TYPE
Syntax
:DISPlay:TYPE <type>
:DISPlay:TYPE?
Description
Set or query the display mode of the waveform on the screen.
Parameter
Name
Type
Range
Default
<type>
Discrete
{VECTors|DOTS}
VECTors
Explanation
VECTors: the sample points are connected by lines. Normally, this mode can provide
the most vivid waveform to view the steep edge of the waveform (such as square
waveform).
DOTS: display the sample points directly. You can directly view each sample point
and use the cursor to measure the X and Y values of the sample point.
Return
Format
The query returns VECT or DOTS.
Example
:DISPlay:TYPE DOTS /*Select dots display mode*/
:DISPlay:TYPE? /*The query returns DOTS*/

Table of Contents

Other manuals for Rigol DS1054Z

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1054Z and is the answer not in the manual?

Rigol DS1054Z Specifications

General IconGeneral
Channels4
Bandwidth50 MHz
Sample Rate1 GSa/s
Memory Depth24 Mpts
Display7 inch TFT LCD (800x480)
Vertical Sensitivity1 mV/div to 10 V/div
Time Base Range5 ns/div to 50 s/div
Vertical Resolution8 bits
Trigger TypesEdge, Pulse, Video, Slope
ConnectivityUSB Host, USB Device, LAN
Input Impedance1 MΩ
Power Supply100-240 VAC, 45-440 Hz