Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-153
:TIMebase Commands
The :TIMebase commands are used to set the horizontal parameters, such as enabling the delayed sweep
and setting the horizontal timebase mode.
Command List:
:TIMebase:DELay:ENABle
:TIMebase:DELay:OFFSet
:TIMebase:DELay:SCALe
:TIMebase[:MAIN]:OFFSet
:TIMebase[:MAIN]:SCALe
:TIMebase:MODE
:TIMebase:DELay:ENABle
Syntax
:TIMebase:DELay:ENABle <bool>
:TIMebase:DELay:ENABle?
Description
Enable or disable the delayed sweep, or query the status of the delayed sweep.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{1|ON}|{0|OFF}}
0|OFF
Explanation
Delayed sweep can be used to enlarge a length of waveform horizontally to view
waveform details.
Return Format
The query returns 1 or 0.
Example
:TIMebase:DELay:ENABle ON /*Enable the delayed sweep*/
:TIMebase:DELay:ENABle? /*The query returns 1*/

Table of Contents

Other manuals for Rigol DS1104Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z Plus and is the answer not in the manual?

Rigol DS1104Z Plus Specifications

General IconGeneral
Bandwidth100 MHz
Channels4
Sample Rate1 GSa/s
Memory Depth24 Mpts
Waveform Capture Rate30, 000 wfms/s
Vertical Sensitivity1 mV/div to 10 V/div
Timebase Range5 ns/div to 50 s/div
ConnectivityUSB Host, USB Device, LAN, AUX
Vertical Resolution8 bits
Power Supply100-240 V AC, 45-440 Hz
Display7 inch TFT LCD
Trigger TypesEdge, Pulse, Video, Slope
Input Impedance1 MΩ ± 2%