EasyManuals Logo

Rigol DS1104Z Plus Programming Guide

Rigol DS1104Z Plus
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #180 background imageLoading...
Page #180 background image
RIGOL Chapter 2 Command System
2-164 MSO1000Z/DS1000Z Programming Guide
:TRIGger:PULSe
Command List:
:TRIGger:PULSe:SOURce
:TRIGger:PULSe:WHEN
:TRIGger:PULSe:WIDTh
:TRIGger:PULSe:UWIDth
:TRIGger:PULSe:LWIDth
:TRIGger:PULSe:LEVel
:TRIGger:PULSe:SOURce
Syntax
:TRIGger:PULSe:SOURce <source>
:TRIGger:PULSe:SOURce?
Description
Set or query the trigger source in pulse width trigger.
Parameter
Name
Type
Range
Default
<source>
Discrete
{D0|D1|D2|D3|D4|D5|D6|D7|D8|
D9|D10|D11|D12|D13|D14|D15|
CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
Explanation
When one or more of channels D0 to D7 is turned on, CH4 cannot be selected; when one
or more of D8 to D15 is turned on, CH3 cannot be selected.
Return
Format
The query returns D0, D1, D2, D3, D4, D5, D6, D7, D8, D9, D10, D11, D12, D13, D14,
D15, CHAN1, CHAN2, CHAN3, or CHAN4.
Example
:TRIGger:PULSe:SOURce CHANnel1 /*Set the trigger source to CH1*/
:TRIGger:PULSe:SOURce? /*The query returns CHAN1*/

Table of Contents

Other manuals for Rigol DS1104Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z Plus and is the answer not in the manual?

Rigol DS1104Z Plus Specifications

General IconGeneral
Bandwidth100 MHz
Channels4
Sample Rate1 GSa/s
Memory Depth24 Mpts
Waveform Capture Rate30, 000 wfms/s
Vertical Sensitivity1 mV/div to 10 V/div
Timebase Range5 ns/div to 50 s/div
ConnectivityUSB Host, USB Device, LAN, AUX
Vertical Resolution8 bits
Power Supply100-240 V AC, 45-440 Hz
Display7 inch TFT LCD
Trigger TypesEdge, Pulse, Video, Slope
Input Impedance1 MΩ ± 2%