Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-3
:SINGle
Syntax
:SINGle
Description
Set the oscilloscope to the single trigger mode. This command is equivalent to any of the
following two operations: pressing the SINGLE key at the front panel and sending
the :TRIGger:SWEep SINGle command.
Explanation
In the single trigger mode, the oscilloscope triggers once when the trigger conditions
are met and then stops.
When the waveform record function is enabled or during the playback of the
recorded waveform, this command is invalid.
Related
Commands
:TFORce
:RUN
:STOP
:TFORce
Syntax
:TFORce
Description
Generate a trigger signal forcefully. This command is only applicable to the normal and
single trigger modes (see the :TRIGger:SWEep command) and is equivalent to pressing
the FORCE key in the trigger control area at the front panel.

Table of Contents

Other manuals for Rigol DS1104Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z Plus and is the answer not in the manual?

Rigol DS1104Z Plus Specifications

General IconGeneral
Bandwidth100 MHz
Channels4
Sample Rate1 GSa/s
Memory Depth24 Mpts
Waveform Capture Rate30, 000 wfms/s
Vertical Sensitivity1 mV/div to 10 V/div
Timebase Range5 ns/div to 50 s/div
ConnectivityUSB Host, USB Device, LAN, AUX
Vertical Resolution8 bits
Power Supply100-240 V AC, 45-440 Hz
Display7 inch TFT LCD
Trigger TypesEdge, Pulse, Video, Slope
Input Impedance1 MΩ ± 2%