Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-7
:ACQuire:SRATe?
Syntax
:ACQuire:SRATe?
Description
Query the current sample rate. The default unit is Sa/s.
Explanation
Sample rate is the sample frequency of the oscilloscope, namely the waveform points
sampled per second.
The following equation describes the relationship among memory depth, sample
rate, and waveform length:
Memory Depth = Sample Rate x Waveform Length
Wherein, the Memory Depth can be set using the :ACQuire:MDEPth command, and
the Waveform Length is the product of the horizontal timebase (set by
the :TIMebase[:MAIN]:SCALe command) times the number of the horizontal scales
(12 for MSO1000Z/DS1000Z).
Return
Format
The query returns the sample rate in scientific notation.
Example
:ACQuire:SRATe? /*The query returns 2.000000e+09*/

Table of Contents

Other manuals for Rigol DS1104Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z Plus and is the answer not in the manual?

Rigol DS1104Z Plus Specifications

General IconGeneral
Bandwidth100 MHz
Channels4
Sample Rate1 GSa/s
Memory Depth24 Mpts
Waveform Capture Rate30, 000 wfms/s
Vertical Sensitivity1 mV/div to 10 V/div
Timebase Range5 ns/div to 50 s/div
ConnectivityUSB Host, USB Device, LAN, AUX
Vertical Resolution8 bits
Power Supply100-240 V AC, 45-440 Hz
Display7 inch TFT LCD
Trigger TypesEdge, Pulse, Video, Slope