EasyManua.ls Logo

Rigol DS1104Z Plus

Rigol DS1104Z Plus
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL Chapter 2 Command System
2-8 MSO1000Z/DS1000Z Programming Guide
:CALibrate Commands
Command List:
:CALibrate:QUIT
:CALibrate:STARt
:CALibrate:QUIT
Syntax
:CALibrate:QUIT
Description
Exit the self-calibration at any time.
Related
Command
:CALibrate:STARt
:CALibrate:STARt
Syntax
:CALibrate:STARt
Description
The oscilloscope starts to execute self-calibration.
Explanation
The self-calibration operation can make the oscilloscope quickly reach its optimum
working state to obtain the most accurate measurement values.
During the self-calibration, all the channels of the oscilloscope must be disconnected
from the inputs.
The functions of most of the keys are disabled during the self-calibration. You can
send the :CALibrate:QUIT command to quit the self-calibration.

Table of Contents

Other manuals for Rigol DS1104Z Plus

Related product manuals