EasyManua.ls Logo

Rigol DS1104Z Plus

Rigol DS1104Z Plus
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL Chapter 2 Command System
2-184 MSO1000Z/DS1000Z Programming Guide
:TRIGger:TIMeout:TIMe
Syntax
:TRIGger:TIMeout:TIMe <NR3>
:TRIGger:TIMeout:TIMe?
Description
Set or query the timeout time in timeout trigger. The default unit is s.
Parameter
Name
Type
Range
Default
<NR3>
Real
16ns to 10s
16ns
Return
Format
The query returns the timeout time in scientific notation.
Example
:TRIGger:TIMeout:TIMe 0.002 /*Set the timeout time to 2ms*/
:TRIGger:TIMeout:TIMe? /*The query returns 2.000000e-03*/