Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-31
:CURSor:AUTO:BY?
Syntax
:CURSor:AUTO:BY?
Description
Query the vertical position of cursor B in auto cursor measurement.
Explanation
The horizontal and vertical positions of the cursor are defined by the pixel
coordinate of the screen. The pixel coordinate of the screen ranges from (0,0) to
(600,400). Wherein, (0,0) is located at the left top corner of the screen and
(600,400) is located at the right bottom corner of the screen. The horizontal pixel
range is from 0 to 600 and the vertical pixel range is from 0 to 400.
In auto cursor measurement, cursor B of Y type is not required for some of the
measurement items. At this point, the query returns 4294967295.
Return
Format
The query returns an integer between 5 and 394.
:CURSor:AUTO:AXValue?
Syntax
:CURSor:AUTO:AXValue?
Description
Query the X value of cursor A in auto cursor measurement. The unit depends on the
horizontal unit currently selected.
Explanation
In auto cursor measurement, cursor A of X type is not required for some of the
measurement items. At this point, the query returns 9.9E37.
Return
Format
The query returns the X value of cursor A in scientific notation.
Example
:CURSor:AUTO:AXValue? /*The query returns -4.000000e-06*/
:CURSor:AUTO:AYValue?
Syntax
:CURSor:AUTO:AYValue?
Description
Query the Y value of cursor A in auto cursor measurement. The unit depends on the
vertical unit currently selected.
Explanation
In auto cursor measurement, cursor A of Y type is not required for some of the
measurement items. At this point, the query returns 9.9E37.
Return
Format
The query returns the Y value of cursor A in scientific notation.
Example
:CURSor:AUTO:AYValue? /*The query returns -4.000000e-01*/
:CURSor:AUTO:BXValue?
Syntax
:CURSor:AUTO:BXValue?
Description
Query the X value of cursor B in auto cursor measurement. The unit depends on the
horizontal unit currently selected.
Explanation
In auto cursor measurement, cursor B of X type is not required for some of the
measurement items. At this point, the query returns 9.9E37.
Return
Format
The query returns the X value of cursor B in scientific notation.
Example
:CURSor:AUTO:BXValue? /*The query returns -4.000000e-06*/

Table of Contents

Other manuals for Rigol DS1104Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z Plus and is the answer not in the manual?

Rigol DS1104Z Plus Specifications

General IconGeneral
Bandwidth100 MHz
Channels4
Sample Rate1 GSa/s
Memory Depth24 Mpts
Waveform Capture Rate30, 000 wfms/s
Vertical Sensitivity1 mV/div to 10 V/div
Timebase Range5 ns/div to 50 s/div
ConnectivityUSB Host, USB Device, LAN, AUX
Vertical Resolution8 bits
Power Supply100-240 V AC, 45-440 Hz
Display7 inch TFT LCD
Trigger TypesEdge, Pulse, Video, Slope
Input Impedance1 MΩ ± 2%