RIGOL Front Panel Key Reference
2-96 User’s Guide for DSA1000 Series
2.6.1.7 Front Switch
Sets the status of the front switch, the default is On. If On is selected, you should
press the power key at the front panel to start up the analyzer after turning on the
rear switch, otherwise the analyzer will start after you turn on the power switch at the
rear panel.
2.6.1.8 Coupl Param
Sets all related parameters automatically according to the coupling relationship.
Definitions of the coupling parameters:
1. CF Step
This parameter maintains a coupling relationship with the RBW (or span) in zero
span (or non-zero span) mode. Refer to CF Step” for more details.
2. Reference level
This parameter maintains a coupling relationship on the max mixing level with
either the input attenuator or the preamplifier. Refer to Ref Levelfor more
details.
3. Input Attenuation
This parameter maintains a coupling relationship on the max mixing level with
either the reference level or the preamplifier. Refer to Ref Levelfor more
details.
4. RBW
This parameter maintains a coupling relationship with the span. Refer to RBW
for more details.
5. VBW
This parameter maintains a coupling relationship with the RBW. Refer to VBW
for more details.

Table of Contents

Other manuals for Rigol DSA1030A series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DSA1030A series and is the answer not in the manual?

Rigol DSA1030A series Specifications

General IconGeneral
BrandRigol
ModelDSA1030A series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals