RIGOL  Front Panel Key Reference 
2-44  User’s Guide for DSA1000 Series 
Key Points: 
  This parameter only changes the readout of the TG output power, rather than the 
actual value of it.  
  The offset could be either a positive (gain) or a negative (loss).   
  You can use the numeric keys, knob or direction keys to modify this parameter. 
For more details, please refer to “Parameter Input”.  
 
Table 2-25 TG Level Offset 
0 dB 
-200 dB to 200 dB 
Unit  dB 
Knob Step  1 dB 
Direction Key Step  10 dB 
 
2.3.4.4  Power Sweep 
Enables or disables the power sweep function. If enabled, the power from the TG 
varies with the sweep rate of the analyzer within the specified range (from start to 
stop frequency) and increases gradually starting from the specified TG power (see 
“TG level”) within specified power range (see “Power Range”). If disabled, the TG 
output is fixed at the specified power (see “TG level”).  
Freq (Hz)
Ampl (dBm)
Start Freq Stop Freq
TG Output
TG Level
Power 
Range
 
Figure 2-14 Power sweep