RIGOL Measurement Examples
3-30 User’s Guide for DSA1000 Series
Figure 3-14 Use marker function to measure TOI distortion
Method 2: (Use TOI function to measure)
Procedures:
1. Device attachment
Connect the output terminals of generators 1&2 to the RF INPUT 50Ω terminal
on the front panel of the analyzer through a power combiner.
2. TOI function to measure
(1) Reset the analyzer
Press Preset
(2) Set the parameters
Press FREQ
Press Center Freq and enter 100 MHz.
Press SPAN
Press Span and enter 1 MHz.
Press Meas
Press Meas Fctn and enter the lower menu.
Page down
Press TOI
(3) Modify the TOI settings
Press Meas Setup