Front Panel Key Reference  RIGOL 
User’s Guide for DSA1000 Series 
2.2  Sweep Settings 
2.2.1  BW 
Sets the RBW (Resolution Bandwidth) and VBW (Video Bandwidth) parameters of the 
analyzer. 
2.2.1.1  RBW 
Sets the desired resolution bandwidth in order to distinguish between signals which 
are close in frequency. 
 
Key Points: 
  Reducing the value of RBW will increase the frequency resolution, but may also 
cause sweeps to take longer (Sweep Time is effected by a combination of RBW 
and VBW when it is in Auto mode). 
  RBW decreases with the span (non-zero span) in Auto RBW mode.  
 
Note: if detector type is “Quasi-Peak” or filter type is “EMI”, RBW can be 200 Hz, 9 kHz, 
120 kHz or 1 MHz only. 
 
Table 2-13 RBW 
Parameter  Explanation 
Default  1 MHz 
Range  100 Hz to 1 MHz 
Unit  GHz, MHz, kHz, Hz 
Knob Step  in 1, 3, 10 sequence 
in 1, 3, 10 sequence 
 
2.2.1.2  VBW 
Sets the desired video bandwidth in order to remove the band noise.