RIGOL  Front Panel Key Reference 
2-32  User’s Guide for DSA1000 Series 
2.2.6  Continue  
Sets the sweep mode to Continue. 
  
Key Points: 
  If the instrument is in Single sweep mode and no measurement function is 
enabled, pressing this key will enter the Continue mode and execute specified 
number of sweeps on the condition that all trigger conditions are satisfied.  
  If the instrument is in Single sweep mode and a measurement function is on, 
pressing this key will enter the Continue sweep mode and execute specified 
number of measurements on the condition that all trigger conditions are satisfied. 
  In Continue mode, the system will send a trigger initialization signal automatically 
and enter the trigger condition judgment directly.  
 
Enter Continue Sweep 
mode and Wait for 
Trigger Condition
Sweep Again and Wait for 
Trigger Condition
Press 
Continue
Single Sweep 
status
?
Y
N
 
 
Figure 2-12 Process of continue sweep