25
NOTE-MH01-EN rev05
o Check the system and verify if alert messages are displayed
o Change the status of MH01 laser system in Ready
o Check the system and verify if alert messages are displayed
o Change the status of MH01 laser system in Standby
o Turn off the system
· After the general check:
o Remove the optical fiber
o Remove the footswitch
o Remove the interlock
o Remove the key
Note: ROCAMED advises wrapping the device with a large quantity of protective plastics.
Note: The shipment of device to the final destination of the customer is under the
responsibility of the distributor. ROCAMED is not responsible for possible damage caused
during this phase.
· Install the device in the room indicated by the customer in the following way:
o Connect the device to the power supply
o Connect the interlock connector
o Connect the footswitch
o Check the laser device
· Perform further controls or additional tests.
· Perform a training to the end user on the following items:
o Proper fiber attachment
o Operation of the device
WARNING:
Do not start any action with the laser device before the official personnel have
performed the installation procedure. The warranty is not comprehensive of any
damage to the laser device before the installation.
4.2.6 Main device connection.
Once all the checks are made and after placing the laser device in its final position in the
working area, you can connect the device to the mains. Use the cable provided. Such cable
can suffer wear over time. The operator or anyone involved in the ordinary maintenance of
the device after the installation must take care of monitoring the state of maintenance of
the power cable.
The device must be connected to the mains in compliance with electrical safety regulations.
The laser system can be connected to a power socket that provides 100 120 VAC, 50-60 Hz,
16A or 200-240 VAC, 50-60 Hz, 10A.
WARNING:
During its use the device must be placed in a position such to keep the power/mains
switch easily accessible.
The power/mains switch is the mean of disconnection from mains.