53
NOTE-MH01-EN rev05
7 MAINTENANCE, CLEANING AND STERILIZATION
This section is not a part of a service or technical assistance manual. It contains only the user
information regarding the care and cleaning of the device, and the cleaning and sterilization
information of optical fiber. For any issues not relating the maintenance contact ROCAMED
Warning:
Refer only to the Manufacturer or authorised distributors for service.
Only trained and qualified personnel are permitted to provide service for this
LASER system. Service by non-authorized people will cause warranty to lapse
7.1 Device cleaning
The laser device MH01 doesn’t need special maintenance by the user. Clean the visible
surfaces with a damp cloth and a caustic solution such as foam. Do not use alcohol or
disinfectant solution, because they are highly flammable. During cleaning be careful not to
let the cleaning solution inside the fiber connection port. Use the supplied caps to close the
fiber connection port after each use. Do not use any alcohol solution to clean the display.
7.2 Laser Maintenance and technical check
Laser device MH01 is designed for maximum safety and performance. With careful use and
under normal operating conditions, the manufacturer recommends a review of the device
by a qualified technician every 12 months.
The intensive use, dust, or continuous movement of the laser in different places may
require a more frequent review.
Warning:
The laser device has to be closed. Only personnel authorized by the
manufacturer can open the external covering panels.
Only personnel authorized by the manufacturer providing technical service
can have access to the internal components of the system.
Only authorized service personnel can replace the replace the power supply
cord
7.3 Safety labelling check
Verifying the integrity and readability of the security labels placed on the device by the
user. I labels are damaged, they must be replaced immediately in accordance with the
plan shown in paragraph labelled.
7.4 Water cooling system
If the level of the water is too low it is necessary to refill the water tank. It is recommended
to check the water status at least every 6 months.