EasyManua.ls Logo

ROCAMED MH01 - Alarms and Warning Description

ROCAMED MH01
71 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44
NOTE-MH01-EN rev05
For all fiber dimension*
Tissue Resection/Cutting
200
mJ
300
mJ
400
mJ
500
mJ
600
mJ
700
mJ
800
mJ
1 J
1.2 J
1.5 J
1.8 J
2 J
2.5 J
3 J
3.5 J
3 Hz
5 Hz
8 Hz
4
W
4,8
W
5,6
W
6,4
W
8
W
9,6
W
12
W
14,4
W
16
W
20
W
24
W
10 Hz
5 W
6 W
7 W
8 W
10 W
12 W
15 W
18 W
20 W
25 W
30 W
12 Hz
6
W
7,2
W
8,4
W
9,6
W
12
W
14,4
W
18
W
21,6
W
24
W
30
W
15 Hz
7,5
W
9
W
10,5
W
12
W
15
W
18
W
22,5
W
20 Hz
10 W
25 Hz
For all fiber dimension*
Tissue - Coagulation
200
mJ
300
mJ
400
mJ
500
mJ
600
mJ
700
mJ
800
mJ
1 J
1.2 J
1.5 J
1.8 J
2 J
2.5 J
3 J
3.5 J
3 Hz
5 Hz
2,5
W
3
W
3,5
W
4
W
5
W
6
W
7,5
W
9
W
10
W
12,5
W
15
W
8 Hz
4
W
4,8
W
5,6
W
6,4
W
8
W
9,6
W
12
W
14,4
W
16
W
20
W
10 Hz
5 W
6 W
7 W
8 W
10 W
12 W
15 W
18 W
20 W
25 W
12 Hz
6
W
7,2
W
8,4
W
9,6
W
12
W
14,4
W
18
W
21,6
W
24
W
15 Hz
7,5
W
9
W
10,5
W
12
W
15
W
18
W
22,5
W
20 Hz
10
W
25 Hz
* fibers with core diameter ≤ 272 µm could be limited (power output) depending to the
technical specification of the connected device.
5.6 Alarms and warning description
There are different alarms that can be displayed by the control panel. Every time that there is
an alarm a yellow screen appears with an error message text that indicates the alarm type.
All the alarms cause the laser system stop: the power electronics are cut off through a suitable
switch and the device is put in a safe mode or is restarted.
The possible types of alarm are the following (see picture below):
· Remote Interlock alarm: the remote connector near the footswitch is not connected or is
connected in a wrong way. This alarm is displayed in the READY mode.
· Footswitch (Pedal) alarm: the footswitch connector is not connected or is connected in a
wrong way. This alarm is displayed in the READY mode
· Shutter alarm: the shutter doesn’t work well. Call Service
· Fiber (Optical Interlock) alarm: the optical fiber is not connected or it’s connected in a
wrong way. Try to unplug and plug again the fiber in the proper way.
· Fiber expired: you are searching to use the connected fiber beyond the expected uses.

Table of Contents