8 TROUBLESHOOTING
Laser devices are designed to have the best performance and maximum safety. If the system
does not work properly, the following table diagnostics can assist in identifying the cause.
Trouble
Possible cause
Solution
Key switch does nothing
· Line power not connected
·
Main line breaker off
·
Emergency red push button
pressed
· Check line cord breaker
·
Push the emergency red
push button
Lamp error
· Simmering error
· Lamp broken
· Lamp old
· Call service
Flow error
· Not enough water
· Water tube twisted
· Fill water reservoir
· Check water tubing
Temperature error
· Room temperature too much high
or air inlet/outlet not free
· Check cooling air circulation
·
Assure the correct distance
of the rear panel from the
wall
Shutter error
Incorrect shutter position
· Call service
Fiber error
Wrong installation of the fiber
· Check the proper
connection of the fiber
Remote interlock error
The ext. interlock is not connected or
is bad connected
· Control the ext. interlock
connection
Footswitch error
The footswitch is not connected or is
bad connected
· Control the footswitch
connection
Energy warning
Fluctuations of the energy
· Wait some second until
the energy is stabilized
·
If the energy always
fluctuates, Call service
No correspondence
between displayed
energy and treatment
effects
Damage of the fiber
· Change fiber
·
Call service