7.7 Use, Cleaning, disinfection, sterilization of optical fibers
Carefully read and follow
Optical fibers instruction manual
7.8 Blast shield replacement
WARNING:
This operation must be done with clean hands and extreme caution to
avoid optical damage of the device.
The blast shield replacement must be executed in the following way:
1. Turn the device off, disconnecting the
equipment from mains
2. Open the Upper cover with the dedicated key
to access to the blast shield area
3. Remove the blast shield unscrewing the knob
4. Check the protective glass for any visible damage
5. In case of damage, substitute the whole blast shield paying attention not to touch the
protective glass with hands.
6. Close the protective door screwing well the doorknob.
7. In case of assistance, please contact Service Department