45
NOTE-MH01-EN rev05
· Water flow alarm (H
2
O pump): the cooling water flow is too low. Check the cooling water
flows freely in the circuit. Check if tubes are truncated or twisted. If water level is low, fill the
reservoir with distilled water only.
· Water temp alarm (H
2
O pump): the temperature of the cooling water is too high. Call
service
· Flash lamp (Simmer) alarm: the flash lamp doesn’t ignite. Call service.
· Energy warning: this warning appears every time there are energy fluctuations and it
stops when the energy value is stable.
Buzzer
The device is equipped with an internal buzzer that emits an acoustic signal with a fixed
duration. It is active in the following case:
· In READY mode when the footswitch is pressed;
· During the operation of the system the signal is emitted synchronously with the laser
pulses.
5.7 Shutdown procedure and protection against unauthorized use
Once finished using the device and the laser is in standby mode, you can proceed with its
shutdown as follows:
· Disconnect the optical fiber from the device and cover with the appropriate plug the laser
device output
· Turn the key switch to position and remove the key to prevent unauthorized use.
· Turn off the main switch on the rear panel and disconnect the power cable
· Disconnect the remote interlock
· Disconnect the footswitch
· Keep the device and accessories in a dry and safe place.