12 TABLES (EMC)
CAUTION!
To guarantee the safety of the user, the patient and others, use only accessories
and spare parts specified by the manufacturer of this product.
Other accessories or spare parts can cause the emission of increased
electromagnetic radiation or reduced immunity against interference.
IMPORTANT!
Medical electrical devices are subject to special precautions with regard to
electromagnetic compatibility (EMC) according to IEC 60601-1-2.
Make sure you observe the notes on EMC for installation and operation.
Medical electrical devices can be influenced by mobile HF communication devices
(i.e. mobile phone) and other devices that may be used in conjunction with them
for treatment or diagnosis.
If it is necessary to stack the devices or place them next to each other, and HF
interference is observed, make sure you observe the intended use of the devices.
IEC 60601-1-2 Table 201
Guidance and manufacturer’s declaration – electromagnetic emission
MH01 is intended for use in the electromagnetic environment specified below. The customer or
the end user of the MH01 should assure that it is used in such an environment
Electromagnetic environment - guidance
MH01 uses RF energy only for its internal
function. Therefore its emissions are very low
and are not likely to cause any interference in
nearby electronic equipment
is not suitable for installation in all
buildings including domestic and those directly
connected to the public supply network in low
tension, but only in buildings like the hospital
with dedicated supply system.
Harmonic emission
IEC 61000-3-2
Voltage fluctuation/flicker
emission