NOTE-MH01-EN rev05
IEC 60601-1-2 Table 202
Guidance and manufacturer’s declaration – electromagnetic immunity
MH01 is intended for use in the electromagnetic environment specified below. The customer or
the end user of MH01 should assure that it is used in such an environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000
-4-2
2 to 8
kV contact
2 to15
kV air
Complies
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2
kV for power
supply lines
Complies
Mains power quality
should be that of a typical
hospital environment
Surge
IEC 61000
-4-5
±1
kV line to line
±2
kV line to earth
Complies
Mains power quality
should be that of a typical
hospital environment
Voltage Dips, Short
interruptions and
voltage variations
on power supply
input lines
IEC 61000
-4-11
>95
% Ut for 0,5
cycle
Voltage interruption
> 95% Ut, 5 “
Complies
Mains power quality
should be that of a typical
commercial or hospital
environment.
If the user of MH01
requires continued
operation dur
ing power
mains interr
uptions, it is
recommended that
MH01
be powered from an
Uninterruptible Power
Supply or Battery
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
Not applicable
There aren’t
elements
susceptible to
magnetic fields
-
Note: Ut is the AC mains voltage prior to application of the test level