64
NOTE-MH01-EN rev05
IEC 60601-1-2 Table 206
Recommended separation distances between portable and mobile RF communication equipment
and MH01.
MH01 is intended for use in the electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of
MH01
can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and
MH01
as recommended below, according to the maximum power of
communications equipment.
Separation distance according to frequency of transmitter
Rated maximum
output power of
transmitter
W
150 KHz to 80 MHz
d=1,17*sqrt (P)
m
80 MHz to 800 MHz
d=1,17*sqrt (P)
m
800 MHz to 2,5 GHz
d=2,33*sqrt (P)
m
0,01
0,117
0,117
0,233
0,1
0,370
0,370
0,740
1
1,17
1,17
2,33
10
3,70
3,70
7,40
100
11,7
11,7
23,3
For transmitters rated at maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer
Note 1: at 80 MHz and 800 MHz, the higher frequency range applies
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by
absorption and reflection from structures, objects and people.