Initial Startup • 9
119
9.8 Options for test strips
This menu allows you to select options for handling lot
numbers and specify whether the operator is allowed to
add or delete reagent lots, edit the expiration date and
limit values on the meter (see “Creating a setup pass-
word” on page 126).
1 Touch the desired option to enable it:
■ Reagent Editing
– Allowed: Enables the operator to edit the reagent
information without needing a password
– Password Needed: Enables the operator to edit the
reagent information only after entering a password
■ Strip Lot Verification
– Display only: The lot number is shown on screen but
the operator cannot confirm it or select a different
lot number.
– Confirmation: The operator must confirm the lot
number displayed and can enter alternative lot
numbers manually or via barcode scanner.
– List selection: The operator can also select from a
list of stored lot numbers.
– Scan only: Lot numbers can be verified via barcode
scanner only.
2 Touch to change to the second screen of
options, or touch to exit this menu without
saving any changes.
When using a data management system for configu-
ration, it is possible to fully disable the Reagent Editing
option.
Reagent Options
Allowed
P
assword Needed
Display only
Confi rmation
List selection
Scan only
Reagent Editing
Strip Lot Verifi cation
Not Allowed
12:48 pm
09/18/13