Glucose Control Testing • 4
59
If a glucose control test is required (as shown in the illus-
tration on the left), you will not be able to test blood glu-
cose until the controls have been run successfully. For
emergencies, STAT tests can be configured in the meter.
This option allows the meter to perform a limited number
of blood glucose tests, if circumstances require, despite
the meter being in QC Lockout (see page 68).
Information stored during
glucose control testing
The following information is stored for every glucose con-
trol test using control solution:
■ Glucose control test result
■ Lot number of the control solution
■ Operator ID (if configured)
■ Level of control solution (L1 or L2)
■ Lot number of the test strips
■ Time and date of test
■ Comments (if applicable)
■ Out of range measurements