Linearity Testing • 7
95
Applying a linearity test sample Once the meter has checked the test strip, the hourglass
icon disappears and you are prompted to apply linearity
solution.
1 Wait until the flashing drop appears in the display
before applying the solution. The meter beeps.
2 Apply a drop of the linearity solution to the front
edge of the test strip. Do not apply the solution to
the top of the strip.
The linearity solution is pulled into the test strip
area by capillary action.
Once sufficient linearity solution has been detected, the
meter beeps and the measurement begins.
Linearity Test
Linearity 12345678 L1
Strip L
ot 545794
12:48 pm
09/18/13
When applying the linearity solution, position the
meter so that the test strip port is always higher or on
the same level as the linearity solution. This prevents
any excess solution from flowing down the strip and
entering the meter.