3 Patient Glucose Testing
3.1 Information regarding blood glucose testing
Preparing to test The following requirements must be met before you can
perform a test:
■ The Accu-Chek Inform II test strips are available.
■ The code file for the test strip lot in use must be
stored in the meter (see Chapter 6).
■ An operator ID must be entered (with password, if
required), if the meter is configured for login.
■ Glucose control tests specified in the system con-
figuration must be run and completed successfully
before running patient tests. Information whether
glucose control tests are necessary is displayed in
the Power Up screen.
■ If configured, the meter may require downloading
stored data to the data management system within
defined time intervals. If such a download (either
via WLAN or by docking the meter in a base unit)
does not happen within the specified time, the
meter is locked (Download Lockout) and cannot be
used for testing.
Contrast
Power Up
QC Due: Immediately
12:48 pm
09/18/13