A • Appendix
162
Measurement Flow
Comments required 0: optional
1: out of range
2: required
3: disabled
0NY
Comments requirement level: if out of…
(only valid if Comments required = 1)
0: normal range
1: critical range
2: reportable range
3: measurement range
0NY
Custom comments 0: enabled
1: disabled
0NY
Control lot verification 0: display only
1: yes / no confirmation
2: prompt for entry
3: scan only
1YY
Control result display 0: value (numeric)
1: PASS / FAIL
0NY
Critical range display 0: value (numeric)
1: HI / LO
0NY
Critical range HI limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
YY
Critical range LO limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
YY
Critical range message enabled Whether to display the
out-of-critical range
warning message (1) or
not (0)
1NY
Critical range text 0 – 100 characters “” N Y
Normal range HI limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
YY
Normal range LO limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
YY
STAT tests allowed 0: no
1: yes
1YY
Number of STAT tests allowed 0 – 9 9 Y Y
Reportable range HI limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
YY
Reportable range LO limit 10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
YY
Reportable range message Whether to display a
reportable range
message (1) or not (0)
1NY
Reportable range message text 0 – 100 characters “” N Y
Subject/Attribute Range Default Device DMS