QC algorithm
QC Algorithm 0: None (always OK)
1: Last result OK
2: Time of day (DMS only)
3: Shift (DMS only)
4: Hours
5: Strip count
6: Time of day rotate
(DMS only)
7: Shift rotate (DMS only)
8: Hours rotate (DMS
only)
9: Strip count rotate
(DMS only)
0 (Y) Y
QC hours (Number of hours between
required control measurements if QC algo-
rithm is set to “Hours” or “hours rotate”)
0 – 9,999 hours 24 Y Y
QC shift length (a PASS control remains
valid for twice this amount in time after
shift start if the QC algorithm is set to
“Shift” or “Shift rotate”)
1, 2, 3, 4, 6, 8, 12, 24 hours 8 N Y
QC shift start (time of the shift start if the
QC algorithm is set to “Shift” or “Shift
rotate”)
0 – 23 hours 6 N Y
QC strip count (number of tests a PASS
control remains valid)
1 – 999 50 Y Y
QC time of day 6 POCT1-A data fields
00:00 – 23:59
06:00
09:00
12:00
15:00
18:00
22:00
NY
QC time of day set 6 POCT1-A data fields
0 or 1
1 0 0 0 0 0 N Y
Subject/Attribute Range Default Device DMS

Table of Contents

Other manuals for Roche ACCU-CHEK Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche ACCU-CHEK Inform II and is the answer not in the manual?

Roche ACCU-CHEK Inform II Specifications

General IconGeneral
BrandRoche
ModelACCU-CHEK Inform II
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals