EasyManua.ls Logo

Rohde & Schwarz CMU200 - Replacing the Option OCXO REFERENCE OSC. CMU-B11 or CMU-B12; Installing the New OPTION OCXO REFERENCE OSC. and Completing the; Manual OCXO Adjustment

Rohde & Schwarz CMU200
216 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MODULE Replacement CMU
1100.4903.82 3.44 E-4
Replacing the Option OCXO REFERENCE OSC. CMU-B11 or CMU-
B12
(see chapter 5, spare part list, Current No. 690 (OCXO REFERENCE OSC. CMU-B11), 710 (OCXO
REFERENCE OSC. CMU-B12
) and explosion drawing 1100.0008.01 D sheet 2)
The Option OCXO REFERENCE OSC. is fitted at the right side panel below the instrument fan in the
CMU. For replacement proceed as follows:
Opening the instrument and removing the OPTION OCXO REFERENCE OSC.
½ Switch off the instrument, pull the mains plug and unscrew the rear-panel feet.
½ Place the instrument onto the front carrying handles, push the tube upwards and take off.
½ Place the instrument onto the left side, locate the OPTION OCXO (see chapter 5, explosion drawing
1100.0008.01 D sheet 2) and unscrew the two combi screws on the module support and one combi
screw on the MOTHERBOARD1.
½ Pull off OPTION OCXO REFERENCE OSC. from the MOTHERBOARD1 connector, slightly tilt and
swing out towards the bottom.
Installing the new OPTION OCXO REFERENCE OSC. and completing the instrument
½ Install the new OPTION OCXO REFERENCE OSC. in the reverse order. Tighten the three combi
screws again.
Note: Make sure that the 10-pin connector to the MOTHERBOARD is inserted correctly.
½ Complete the instrument without causing damage to the cables.
Manual OCXO adjustment
½ Connect the instrument to the mains and switch on.
½ Check the frequency accuracy and, if necessary, perform manual OCXO adjustment as described in
chapter 2.