New Installation of the CMU Software CMU
1100.4903.82 4.2 E-4
Performing a Software Update
On start-up, an automatic test is made to check whether an update floppy disk or a flash disk/hard disk
is available. If this is the case, the version manager is started, displaying the following possible
selections:
VersionManager Ver X.X
The active CMU software is the version: [active version]
activate other software version
delete a software version
install software version from FLOPPY DISK
DRIVE or PCMCIA Slot X
firmware update after board change
edit service tables
exit
In the above menu, an action can be selected using the softkey located to the left of the arrow.
½ For the software update select the action install new software version from the Floppy DISK DRIVE
or the PCMCIA Slot X.
½ Press the softkey to the left of the above mentioned menu item. An overview of the software versions
that already provided on the installation media will appear.
½ Select the desired version by pressing the softkey to the left of the arrow.
The software update will then start automatically, prompting the user to insert a further update floppy
disk or flash disk, if necessary. Confirm using the ENTER hardkey.
Note: In the case of several update floppy disks or flash disks always observe the order of the
disk numbers.
½ When all update floppy disks or flash disks have been loaded, follow the prompt
Remove installation media and press ENTER key
The CMU boots again and automatically performs a firmware update of the hardware boards. This may
take a few minutes. Then the new software version will start and the CMU is ready for use.