3 Modification of Universal Link Handler
Remove the two screen covers after undoing the screws (Fig. 4).
If the TAZ (update state) of the Link Handler base plate (1097.3570.02) is lower than -03.06-, modify as
shown in Fig. 5.
The update status (TAZ) can be seen on the barcode label stuck to the solder side of the board inside the
screening strip.
If the TAZ of the Link Handler base plate in slot A430 (1097.3570.02) is lower than -03.06-, modify the top
screen cover (labelled) as shown in Fig. 6.
If the update status (TAZ) of the I/Q Out module on slot A430 (1100.2000.02) is lower than –03.08-, modify
as shown in Fig. 7.
The update status (TAZ) is visible when the covers are removed.
If the optional Speech Codec CMU-B52 (1100.5400.02) is fitted, remove the module in slot A420 (it is no
longer required, Fig. 4).
Insert Bluetooth module 1135.6502.02 (Fig. 4).
Stick the OPTION INCL. (1100.5797.00) label to the cover at the solder side (Fig. 4).
Screw back the two screen covers and insert Link Handler into the unit (Fig.3 and 4).