R&S PB Installation Instruction CMU-U53 BLUETOOTH_F.CMUBi/1CMK 08.09.00 1100.7383.00 01.00
page 10 of 11
4 Assembling the Unit
Fit cover and cabinet feet back on in the reverse order to that described under 1.2/1.3/1.4 (Fig. 1 and 2).
5 Functional Test of Modification Kit CMU-U53
Connect unit to AC supply and switch on.
Automatic module data adjustment
During startup observe the display.
When three BEEPs can be heard, press the MENU SELECT key.
The VersionManager is started (see also chapter 4 in CMU 200 manual, Firmware Update). The display
includes the menu item FIRMWARE UPDATE AFTER BOARD CHANGE.
Press the softkey to the left of the above mentioned menu item.
The automatic module data adjustment is performed under software control, firmware updates for
microprocessors and programmable devices being performed as well. This may take a few minutes.
After the adjustment has been terminated, the operating software starts automatically and the CMU is ready for
use and complies with the specifications.
Test of Option Detection
Software version > 2.60 must be provided in the instrument.
Press SETUP hardkey in SYSTEM panel.
Press OPTIONS softkey located on the row of keys at bottom of display.
--> Check whether CMU-B21 and CMU-B53 is active.
Install software option CMU-K53 and activate key code (see instructions supplied with the modification kit).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz CMU200 and is the answer not in the manual?

Rohde & Schwarz CMU200 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelCMU200
CategoryTest Equipment
LanguageEnglish

Related product manuals