R&S PB Installation Instruction CMU-U53 BLUETOOTH_F.CMUBi/1CMK 08.09.00 1100.7383.00 01.00
page 11 of 11
Calibration
Since the Rx/Tx module was replaced, a complete instrument calibration (ACS) has to be performed:
Calibrate the FM path in the Maintenance menu.
Bluetooth Functional Test
Connect a Bluetooth module to RF2 via RF cable or antenna.
Perform inquiry procedure.
Set up a connection to the Bluetooth DUT.
Check results in Power and Modulation menu.