EasyManua.ls Logo

Rohde & Schwarz CMU200 - Preparing the Instrument

Rohde & Schwarz CMU200
216 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Adjustment CMU
1100.4903.82 2.2 E-4
Preparing the Instrument
Opening the casing:
Note:
½ Remove the power plug on the CMU and place the CMU onto the
front handles.
½ Loosen the four Phillips screws at the four rear-panel feet and
take off the feet.
½ Pull off the instrument tube towards the top.
½ Locate the adjustment devices (see chapter 5, drawing
1100.0008.01 D page 3).
½ After performing the manual adjustment, close the casing again
in the reverse order.
If only the Option OCXO REFERENCE OSC. CMU-B11/B12 is to be
adjusted, it is not necessary to open the complete tube of the CMU.
The adjustment can be made from outside through the ventilation
holes using a small screwdriver (see chapter 5, drawing
1100.0008.01 D page 3).
Adjusting the + 5 VDC Reference Voltage
Preparation: ½ Connect the DC voltmeter test cable to the SMB plug X221 (see
chapter 5, drawing 1100.1027.01 page 1) on the motherboard.
½ Switch on the CMU and allow it to warm up for approx. 5 minutes.
Adjustment: ½ Use potentiometer R120 on the REFERENCE BOARD to adjust
the measured value at the DC voltmeter to + 5.000 V ± 1 mV.

Table of Contents

Related product manuals