CMU Troubleshooting
1100.4903.82 3.59 E-4
•  Example: Info menu REFERENCE BOARD
Fault: The frequency accuracy of the RF signal deviates.
Call the Maintenance Menu via the INFO Key select REFERENCE BOARD, start the diagnosis
measurement using the TEST Key and activate the ON hardkey.
All diagnostic voltages 
within permissible 
limits (deviations 
marked red)?
Yes
Connect ext. 
reference source with 
correct frequency!
Ext. reference source 
selected, but not 
connected to REFIN?
Int. reference element TCXO or 
OCXO, if present, incorrectly 
adjusted, faulty or REFERENCE 
BOARD faulty!
No
No
 Int. reference element TXCO or 
OCXO, if present, incorrectly 
adjusted or faulty!
Ext. reference element 
connected, but wrong 
frequency selected!
External reference 
source selected?
No
Yes
Yes