CMU Test Report
1100.4903.82 1.19 E–4
Item
No.
Description
Measure-
ment to
section
Min. Actual Max.
Measurement
uncertainty
Unit
REF OUT 2 13 MHz or 10 MHz REF OUT 2 1 5 V(pp)
REF OUT 2 13 MHz or 10 MHz –1+1Hz
TX Frequency Accuracy
TX Frequency accuracy 2200 MHz TX
Frequency
accuracy
–2200
–220
–11
+2200
(TCXO),
+220
(B11),
+11
(B12)
Hz
TX Frequency Settling
TX frequency settling time
F1 = 100 MHz ->F2 = 200 MHz to
<1 kHz
TX
Frequency
settling
400
µs
TX frequency settling time
F1 = 1800 MHz ->F2 = 1900 MHz to
<1 kHz
400
µs
TX frequency settling time
F1 = 2200 MHz ->F2 = 2100 MHz to
<1 kHz
400
µs
TX frequency settling time
F1 = 100 MHz ->F2 = 2200 MHz to
<1 kHz
400
µs
TX frequency settling time
F1 = 2000 MHz ->F2 = 100 MHz to
<1 kHz
400
µs
VSWR
VSWR RF1 10 MHz VSWR 1.2
VSWR RF1 900 MHz 1.2
VSWR RF1 1800 MHz 1.2
VSWR RF1 2700 MHz 1.6
VSWR RF2 10 MHz 1.2
VSWR RF2 900 MHz 1.2
VSWR RF2 1800 MHz 1.2
VSWR RF2 2700 MHz 1.6
VSWR RF3 OUT 10 MHz 1.5
VSWR RF3 OUT 900 MHz 1.5
VSWR RF3 OUT 1800 MHz 1.5
VSWR RF3 OUT 2700 MHz 1.7
VSWR RF4 IN 10 MHz 1.5
VSWR RF4 IN 900 MHz 1.5
VSWR RF4 IN 1800 MHz 1.5
VSWR RF4 IN 2700 MHz 1.6