Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
229User Manual 1173.9240.02 ─ 17
SOUR:EXT:FREQ:FACT:DEN 2
SOUR:EXT:FREQ:OFFS 100KHZ
//--------------Performing the calibration measurement -----------
//Perform a transmission measurement with direct connection between the generator
//and the analyzer and wait till the end
SENS:CORR:METH TRAN
SENS:CORR:COLL:ACQ THR; *WAI
//--------------Retrieving the calibration trace results -----------
//Retrieve the measured frequencies (10 Hz - 600 kHz)
TRAC:DATA:X? TRACE1
//Retrieve the measured power levels; = 0 between 10 Hz and 100 kHz (below
//generator minimum frequency); nominal -5dBm as of 100 kHz;
TRAC:DATA? TRACE1
//--------------Normalizing the calibration trace results -----------
//Retrieve the normalized power levels (= power offsets from calibration results)
//Should be 0 for all sweep points directly after calibration
SENS:CORR:STAT ON
TRAC:DATA? TRACE1
//--------------Changing the display of the calibration results -----------
//Shift the reference line so the -5 dB level is displayed in the center
DISP:TRAC:Y:SCAL:RVAL -5DB
DISP:TRAC:Y:SCAL:RPOS 50PCT
11.4.3 Configuring the Output
Configuring trigger input/output is described in chapter 11.4.7.2, "Configuring the Trig-
ger Output", on page 253.
DIAGnostic:SERVice:NSOurce........................................................................................229
OUTPut:ADEMod[:ONLine][:STATe]................................................................................ 230
OUTPut:ADEMod[:ONLine]:SOURce............................................................................... 230
OUTPut:ADEMod[:ONLine]:AF[:CFRequency].................................................................. 230
OUTPut:ADEMod[:ONLine]:PHONes............................................................................... 231
DIAGnostic:SERVice:NSOurce <State>
This command turns the 28 V supply of the BNC connector labeled NOISE SOURCE
CONTROL on the front panel on and off.