Instrument Functions
NRP2
143User Manual 1173.9140.02 ─ 07
to windows", on page 144). The name can therefore be changed and deleted as
desired, but above all it is not automatically deleted when one of the preset parameters
is changed and the instrument setting no longer corresponds to the loaded standard.
A table of sensor parameter values for the different standards can be found in chap-
ter A.3, "Standard Configurations",
on page 368.
Fig. 4-69: Recall Standard dialog box
(Select standard setting)
Selects a mobile radio standard.
Remote command:
SYSTem:STANdard:
CATalog? on page 327
SYSTem:STANdard:PRESet on page 327
Preserve Window Settings
If enabled, the measurement windows are not reconfigured.
Remote command:
SYSTem:STANdard:
PWSettings on page 327
External Trigger
If enabled, an external trigger is required. If disabled, the internal trigger is used..
External triggering should be selected if the test signal yields no clearly defined or only
an unstable trigger event.
Note: If external triggering is used in the timeslot mode, make sure that the trigger
delay setting is such that the start of the first timeslot coincides with the delayed trig-
ger, see also chapter 4.2.6, "Trigger settings", on page 85.
Remote command:
SYSTem:
STANdard:TRIGger:SOURce on page 327
Recall
Loads the selected standard settings.
The same effect can be achieved by pressing the MENU
key.
Remote command:
SYSTem:STANdard:PRESet on page 327
Management of settings