Instrument Functions
NRP2
153User Manual 1173.9140.02 ─ 07
Fig. 4-82: Out1 mode / Analog
Remote command:
OUTPut:RECorder<[1]..2>[:
STATe] on page 258
Window 1 2 3 4 ← Analog ← Mode
Selects the window the output refers to.
Only windows that are operated in "Dig" or "D&A" window mode are available for selec-
tion. See chapter 4.3.3.3, "Selecting the result display mode", on page 93 for the set-
ting the display mode.
Remote command:
OUTPut:RECorder<[1]..2>:FEED on page 255
OUTPut:TTL:FEED on page 259
0 V Equivalent - Analog ← Analog ← Mode
Entry of the measurement value that corresponds to the output voltage of 0 V.
Remote command:
OUTPut:RECorder<[1]..2>:LIMit:LOWer on page 256
OUTPut:RECorder<[1]..2>:LIMit:LOWer:POWer on page 256
OUTPut:RECorder<[1]..2>:LIMit:LOWer:RATio on page 256
3 V Equivalent - Analog ← Analog ← Mode
Entry of the measurement value that corresponds to the output voltage of 3 V.
Remote command:
OUTPut:
RECorder<[1]..2>:
LIMit:UPPer on page 257
OUTPut:RECorder<[1]..2>:LIMit:UPPer:POWer on page 257
OUTPut:RECorder<[1]..2>:LIMit:UPPer:RATio on page 257
Pass/Fail ← Mode
The output provides a logic signal in the selected window to indicate that a limit value
has been exceeded.