Remote Control - Commands
NRP2
283User Manual 1173.9140.02 ─ 07
Parameters:
<rlevel> float_value
Range: depending on sensor
*
RST: depending on sensor
Default unit:
DBM
[SENSe<[1]..4>]:STATistics:TIME < time_interval>
Determines, together with the SENS:STAT:OFFS
command, the time interval in which
the power for the statistical evaluation is measured. SENS:STAT:TIME determines the
length of the interval (see also [SENSe<[1]..4>][:POWer]:TGATe<[1]..4>:
TIME on page 276).
Parameters:
< time_interval> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit:
S
[SENSe<[1]..4>]:TIMing:EXCLude:STARt <exclude_start>
Sets a time that is to be excluded at the beginning of the integration period.
Integration period
EXCLude:STARt EXCLude:STOP
Power
Time
Fig. 6-15: Effect of commands SENS<[1]..4>:TIM:EXCL:STAR and :STOP
Parameters:
<exclude_start> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit:
S
[SENSe<[1]..4>]:TIMing:EXCLude:STOP <exclude_stop>
Sets a time that is to be excluded at the end of the integration period. See fig-
ure 6-15 Effect of commands SENS<[1]..4>:TIM:EXCL:STAR and :STOP.