TRIGger<[1]..4>[:IMMediate]........................................................................................... 333
TRIGger:ALL[:IMMediate]
............................................................................................... 333
TRIGger<[1]..4>:LEVel................................................................................................... 333
TRIGger:ALL:LEVel........................................................................................................333
TRIGger<[1]..4>:MASTer[:STATe]................................................................................... 334
TRIGger<[1]..4>:SLOPe..................................................................................................334
TRIGger:ALL:SLOPe......................................................................................................334
TRIGger<[1]..4>:SOURce............................................................................................... 334
TRIGger:ALL:SOURce................................................................................................... 334
TRIGger<[1]..4>:SYNChronize[:STATe]............................................................................335
ABORt<[1]..4>
ABORt:ALL
Immediately sets the respective sensors to the IDLE state. Measurements in progress
are interrupted. If INIT:CONT ON is set, a new measurement is immediately started
since the trigger system is not influenced.
Usage: Setting only
INITiate<[1]..4>:CONTinuous <state>
INITiate:ALL:CONTinuous
<state>
Selects either single-shot or continuous (free-running) measurement cycles. In the
course of a measurement cycle, a sensor passes through the IDLE, INITIATED,
WAIT_FOR_TRG and MEASURING states before it is again set to the IDLE (INIT:
CONT OFF) or the INITIATED (INIT:CONT ON) state. The settings are not changed
when control is switched from manual to remote. INIT:CONT ON is set when remote
control is switched back to manual control.
Parameters:
<state> ON | OFF
ON
Continuous measurements are performed. If a measurement is
completed, the respective sensors do not return to the IDLE
state but are immediately set to INITIATED and then to
WAIT_FOR_TRG.
OFF
A measurement cycle is only performed once. After completion,
the sensors remain in the IDLE state. INIT:
CONT OFF has no
effect when the sensor is already in the IDLE state. A measure-
ment in progress is completed.
*RST: OFF
Pressing the PRESET key or sending the SYSTem:PRESet
command sets the
INITiate:ALL:CONTinuous command to
ON.