EasyManua.ls Logo

Rohde & Schwarz R&S NRP - Additional Functions; Relative Measurements; Setting the Reference Value Manually

Rohde & Schwarz R&S NRP
350 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S NRP Configuring measurements (Measurement menu)
1144.1400.12 4.39 E-2
Additional functions
Relative measurements
The Rel function is a simple way of obtaining the ratio of a measured value to a constant value, e.g. a
measured power.
Measurement menu:
Rel – ( )
Off On Reset ää
Off:
Measurement using the measurement function that has been set.
Rel –
( )
Off On Reset ää
On: Relative display mode.
Remote Control
CALCulate[1..8]:RELative:STATe[?] ON | OFF
Rel –
( )
Off On Reset ää
If the right-hand side of the rocker switch is pressed when On is se-
lected, the current measured value is transferred to the active window’s
reference memory. The display then returns to
On.
Fernsteuerung
CALCulate[1..8]:RELative[:MAGNitude]:AUTO[?] ONCE
Setting the reference value manually
The reference value can also be directly entered or edited using the Ref. Val… dialog box.
Measurement menu:
Ref. Val… ( )
ä
Fig. 4-33 Ref. Value dialog box
Ref. Value ( )
ä ä
For directly entering a new reference value or for editing an old value.
Remote Control
CALCulate[1..8]:RELative[:MAGNitude][?] <float_value>
Result Ref
( )
ä ä
The current result is transferred to the reference value memory of the
active window and displayed. This function can also be started from
the Measurement menu by using the
Reset softkey.
Remote Control
CALCulate[1..8]:RELative[:MAGNitude]:AUTO[?] ONCE
MeasurementRel
MeasurementRef. Val

Table of Contents